USER MENU DESCRIPTIONS
NAME/CODE
DESCRIPTION
CODE/VALUE
A1
Baud Rate
Selects the baud rate for data transmission through serial port
1200 2400
4800
9600
√
A2
Data Bits and
Parity
Selects the number of data bits and parity of serial transmission.
"8n" = 8 data bits with no parity bit and one stop bit
"7O" = 7 data bits with odd parity bit and one stop bit
"7E" = 7 data bits with even parity bit and one stop bit
"7n" = 7 data bits with no parity bit and two stop bits
8n
√
7O
7E
7n
A3
COM1 Mode
Selects the mode of the COM1 serial port:
"0" = ON Demand
"1" = Continuous
0
√
1
A4
Display Check
Automatically checks all display segments
A5
N/A
A6
N/A
A7
Enter Operator
ID
Allows the user to Enter an operator ID number up to 6 digits. Refer
to A21 to include in print format. Us the left/right arrow keys to select
digit and up/down arrow keys to change value. Press Set key to
enter.
Press Down Arrow
key to begin.
A8
Time Entry
Actuates the function that allows entry of the current time. Use the left
or right arrow keys to select the digit and the up/down arrow keys to
change the value. Save by pressing the SET Key.
Press Down Arrow
key to begin
A9
Time Format
Actuates the operator to set the clock as a 12hr or 24 hr setting.
"0" = 12 hour "1" = 24 hour
0
√
1
A10
Date Entry
Actuates the function that allows entry of the current date. Use the
left or right arrow keys to select the digit and the up/down arrow keys
to change the value. Save by pressing the SET Key.
Press Down Arrow
key to begin
A11
Date Format
Selects the format for current date.
"0" = mm/dd/yy "1" = dd/mm/yy
0
√
1
A12
N/A
.
A13
Set COM Port
Actuates the function that sets either RS-232 or RS-485. Save by
pressing the SET Key
RS-232
√
RS-485
A14
N/A
A15
N/A