Neptune 5500
15
6. If the calibration was
not
successful, one of the error messages below will appear. Take
the indicated action to correct the problem, then perform a new calibration.
"
Err0
" - The calibration test weight or the adjusted keyed-in weight is larger than the full
capacity of the scale. Change the calibration test weight or check the input data.
"
Err1
" - The calibration test weight or the adjusted keyed-in weight is smaller than 1%
of the full capacity of the scale. Change the calibration test weight or check the input
data.
"
Err2
" - The internal resolution of the scale is not high enough to accept the calibration
value. Select a larger parameter for the Span Gain (F2). SEE APPENDIX C FOR
MORE INFORMATION.
5.4
VIEW CALIBRATION VALUES (F18)
Note:
The values displayed in this procedure are valid only after a successful calibration
has been performed using F16 and F17.
1. While in the Setup mode, using the right or left arrow keys, scroll to "
F 18
", then scroll
down once using the down arrow
key to enter view calibration menu.
2. The display will momentarily show "
CAL 0
" followed by a value. This value is the
zero
calibration value
and should be recorded in the table below.
3. Press down arrow again. Scale will momentarily show "
CAL 1
" followed by another
value. This value is the
span calibration value
and should also be recorded in the
table below. Press any key to return to upper level (F18).
INDICATOR
ZERO CALIBRATION VALUE
SPAN CALIBRATION VALUE
S/N:
Table 6-1: Calibration Value Table
5.5
KEY-IN ZERO CALIBRATION VALUE (F19) FOR EMERGENCY USE ONLY
Note:
This procedure is intended for emergency use only in the case of non-volatile
memory loss. A valid span calibration value, obtained from a successful F17
calibration procedure, must be used.
1.
While in the Setup mode, using the right or left arrow keys, scroll to "
F 19
", then scroll
down once using the down arrow
key.
2.
The display will momentarily show
"ET CO",
followed by the right digit flashing. Use
the directional keys to adjust the displayed value to the original zero value.
3.
After setting the exact value, press the
SET
key to save the value.
4.
If the entered value is greater than zero, the display will show "
E E CO
" momentarily,
then revert back up to F19. If a value of zero is entered, the indicator will briefly show
"
Err 5
", then revert back to the screen described above in Step # 2.