5.6
KEY-IN CALIBRATION VALUE (F20) FOR EMERGENCY USE ONLY
Note:
This procedure is intended for emergency use only in the case of non-volatile
memory loss. A valid span calibration value, obtained from a successful F17
calibration procedure, must be used.
1.
While in the Setup mode, using the right or left arrow keys, scroll to "
F 20
", then scroll
down once using the down arrow
key.
2.
The display will momentarily show
"ET C1",
followed by the right digit flashing. Use
the directional keys to adjust the displayed value to the original calibration value.
3.
After setting the exact value, press the
SET
key to save the value.
4. If the entered value is greater than zero, the display will show "
E E C1
" momentarily,
then displays “
E T C2”
. Press
SET/GROSS
key and the display will show "
E E C3
"
momentarily, then displays “
E T C3”
. Press
SET/GROSS
key and the scale then
displays "
E E C3
" and goes back to F20.
5.
If a value of zero is entered, the indicator will briefly show "
Err 5
", then revert back to
the screen described above in Step # 2.