Neptune 5500
12
3.3.5
ENTERING A MAX AND MIN VALUE FROM THE KEYBOARD
1. To reset scale to weighing mode without max and min values, press and hold the
PRINT button for three seconds.
2. Make sure the scale is displaying zero weight. If necessary, press the ZERO key.
3. Press the UNIT key until the MIN annunciators light and right digit blinks on the
display.
4. Use the left arrow key to move to the desired digit.
5. Use the up or down arrow keys to increment to the desired digit.
6. Repeat for other digits.
7. Press the SET/GROSS key to enter minimum value.
8. Scale momentarily displays SET then PUT --.
9. Press UNIT key. The MAX annunciators light and right digit blinks on the display.
10. Use the left arrow TARE key to move to the desired digit.
11. Use the up or down arrow keys to increment to the desired digit.
12. Repeat for other digits
13. Press the SET/GROSS key to enter maximum value.
14. The scale is now ready to check weigh.
15. To reset scale to weighing mode without max and min values, press and hold the
PRINT button for three seconds.
3.3.6
ENTERING A MAX AND MIN VALUE FROM KNOWN WEIGHT
1. To reset scale to weighing mode without max and min values, press and hold the
PRINT button for three seconds.
2. Make sure the scale is displaying zero weight. If necessary, press the ZERO key.
3. Place the known minimum weight on the scale platform.
4. Press the UNIT key until the MIN annunciators light and right digit blinks on the
display.
5. Press the SET/GROSS key to enter minimum value
6. Scale momentarily displays SET then PUT --.
7. Place the known maximum weight on the scale platform.
8. Press the UNIT key to enter maximum value
9. Press the SET key to finish.
10. The scale is now ready to check weigh
11. To reset scale to weighing mode without max and min values, press and hold the
PRINT button for three seconds.
3.3.7 ENTERING A SINGLE TARGET VALUE (min and max values are equal)
1. To reset scale to weighing mode without max and min values, press and hold the
PRINT button for three seconds.
2. Make sure the scale is displaying zero weight. If necessary, press the ZERO key
3. Press the UNIT key until the MIN annunciators light and right digit blinks on the
display.
4. Use the left arrow (TARE) key to move to the desired digit
Use the up or down arrow keys to increment to the desired digit.
5. Repeat for other digits
6. Press the SET/GROSS key to enter single target value.
7. Scale momentarily displays SET then PUT --.
8. Press UNIT key. The MAX annunciators light and right digit blinks on the display.
9. Press the SET/GROSS key to enter the single TARGET value.
10. The scale is now ready to check weight. The under lights will be on when the weight
is less than min, the target lights will be on when the weight is equal to min and the
over lights will be on when the weight is greater than the min.