Calibration
RMWH-SVX001C-EN
29
Unsuccessful Calibrations
In the unlikely event that a problem occurs during a
calibration process, the
Results
screen will show the
calibration status as
Unsuccessful
.
1. On the
Results
screen, select
Continue
.
2. On the
Calibration Unsuccessful
screen, do one of
the following:
a. For zero gas calibrations:
i. To retry the zero gas calibration, select
Repeat
Calibration
.
ii. To return to the previous calibration values,
select
Close
.
b. For span gas calibrations:
i. To do a span gas calibration for another gas,
select
Select Another Gas
.
ii. To retry the span gas calibration, select
Repeat
Calibration
.
iii. To return to the previous calibration values,
select
Close
.
If
Close
is selected, on the dashboard, a gray check
mark indentifies the points that are affected by the
unsuccessful calibration. The gas readings for these
sample points are below the alarm threshold but may
not be completely accurate.