Calibration
28
RMWH-SVX001C-EN
10. On the
Span Gas Calibration
screen, select
Start
.
A progress bar shows during the span gas calibration.
11. When span gas calibration is complete, the
Results
screen shows. Select
Continue
.
12. On the
Calibration Complete
screen, do one of the
following:
a. To do a span gas calibration for another gas, select
Select Another Gas
. Repeat
through
b. To return to the
Results
screen, select
Back
.
c. To go to the next step, select
Continue
.
13. Close the span gas cylinder valve.
14. On the
Set Reminder
screen, select the applicable
button for the calibration interval. At the selected
interval, a pop-up will signal the need for calibration.
15. Select
Next
.
16. On the bottom of the device, install the pin in the
calibration port.
17. Remove the tubing from the calibration port.
18. On the device touchscreen, select
Done
.