- 76 -
4.1.6 A0, A1, A2 and A3 Command
With these commands, the measurement mode of the instrument is changed.
This setting is stored although the power is turned OFF. If an
“ST” command is sent without sending
one of these commands, measurement is carried out using the measurement mode set before the
“ST” command was sent.
Command
Measurement mode
A0 :
AUTO
A1_##### :
FREQ
"_" is space. "#####" is frequency. 1.5~250 Hz
A2_##### :
MANU
"_" is space. "#####" is integral time.
10-120000ms :SR-LEDW
20-120000ms :SR-UL2,SR-UL1R
20-15000ms :SR-3A
A3 :
SYNC
PC
Command communication direction
Measuring instrument
"A0", "A1" "A2" or "A3" + (Cr
•Lf)
“OK”+(Cr•Lf)