- 88 -
4.1.30
FG Command
Reads current setting in FIX INTEG mode.
(
SR-LEDW only
)
PC
Command communication direction
Measuring instrument
”FG” + (Cr
・
Lf)
“OK” + (Cr
・
Lf)
Integral time+ (Cr
・
Lf)
Inner- (Cr
・
Lf)
“END” + (Cr
・
Lf)
・
Integral time
:
[ms]unit
・
Inner-filter type
:
1
→
FILTER
1
2
→ FILTER
2
3
→
FILTER
3
4
→ FILTER
4
5
→
FILTER
5
4.1.31 FS Command
Sets inner-filter type and integral time in FIX INTEG mode. (SR-LEDW only
)
The left side of # means integral time [ms](50 -120,000),
right side means inner-filter (1 - 5)
FS_###_#
:
The left side of # means integral time[ms](50 -120,000).
The
right side means inner-filter (1-5)
PC
Command communication direction
Measuring instrument
”FS_Integral time_Filter type” + (Cr
・
Lf)
OK+(Cr
・
Lf)
Remember!
The integral time and inner-filter setting in FIX INTEG mode are erased
when the power off.