13
Instructions for Use
TNI
soft Flow
50
Clinic System
2.1.3 Water Bag installation
• When installing or changing the bag with sterile water the device has to be switched off.
• Place the water bag so that the opening is 1 m above the upper edge of the device (see figure
below as example).
• Push the spike of the chamber hose into the dedicated opening at the bottom of the water bag.
• Open the vent cap on the side of the bag spike. The humidification chamber will now automatically
and constantly be filled up to the mark line until the water bag is empty.
NOTE
•
Ensure that the humidifi cation chamber and the water
bag always contain suffi cient amounts of water.
•
Switch off the TNI softFlow system if not in use.
WARNING
• Ensure that the water level is always between the black
marking lines (see Image)!
• Use sterile water only. Do not use any additives!
•
Empty the humidifi cation chamber completely before
transporting or moving the device!
•
Do not use the humidifi cation chamber if it shows
visible damage!
2.1.4 Components of Humidifi er Homecare
The Humidifi er Homecare consists of three parts.
Lid Humidifi er Homecare
Cyclone Element Humidifi er Homecare
Water Chamber Humidifi er Homecare
Example image