![TNI 40610021 Instructions For Use Manual Download Page 31](http://html.mh-extra.com/html/tni/40610021/40610021_instructions-for-use-manual_1122044031.webp)
31
Instructions for Use
TNI
soft Flow
50
Clinic System
5 Technical information
5.1 Product specifications
Performance data
Flow rate
10 to 50 l/min (adjustable in 0.5 l/min steps)
Admixture of oxygen
0-20 l/min
Humidity dew point
30-37°C DP (adjustable in 1°C DP steps)
Event memory
Data storage of the last 12 therapy months
Device parameters
Technical data
Medical product class (93/42/EWG)
IIa
Safety class, electrically
II
Sound level
< 30 dB(A)
Alarm signal sound pressure
> 60 dB(A)
Safety type
IP21
Applied part (applicator)
BF
Electrical safety
According to EN 60601-1
UL 60601-1
CSA C22.2/No 60601-1
Electromagnetic compatibility
According to EN 60601-1-2
Operating voltage (nominal voltage)
100-240 V
~
, 50-60 Hz
Maximum power system
300 VA
Maximum power applicator heating
20 VA
Device dimensions
Width
320 mm
Depth
320 mm
Height
210 mm
Weight (without humidifier, without water)
5.6 kg
Humidification chamber auto-fill
max. 144 ml
Sterile water bag or bottle
< 1000 ml
Water tank humidifier homecare
max. 650 ml
Applicator
Changing cycle applicator clinic series
Changing cycle applicator homecare series
≤ 360 therapy hours; single-patient use
≤ 720 therapy hours; single-patient use
Safety level (applied part)
BF
Tube length nasal application
1.8 m
Tube length tracheal application
2,33 m
Max. temperature of air leaving device
43°C
Summary of Contents for 40610021
Page 1: ......
Page 2: ...TNI System ...