15
Instructions for Use
TNI
soft Flow
50
Clinic System
2.2 Applicators
NOTE
• To meet the requirements of the ongoing therapy, be sure to choose the appropriate applica-
tor type.
• Keep the heated applicator tube away from any electronic monitoring electrode (EEG, ECG,
EMG, etc.) to avoid potential interference with the monitored signal.
• Do not jam or bend the tube.
• The applicator must be changed with every patient.
WARNING
To avoid the risk of burns:
• Do not use accessories that are not authorized by TNI medical AG.
• Do not use insulating sleeves and do not cover the applicator when in use (e.g. by a blanket).
• Do not use any external source (a radiant heater, e.g.) to heat the applicator.
• Do not modify the applicator in any way.
To avoid the risk of electric shocks:
• After the applicator has been attached, the patient should not touch the electrical connec-
tions of the TNI softFlow system.
2.2.1 Applicator installation
Choose appropriate applicator type.
Insert the applicator plug from above into the dedi-
cated socket and push it down gently and fully.