Instructions for Use
TNI
soft Flow
50
Clinic System
Humidifier
Typical humidity nasal application
Typical humidity tracheal application
30-37°C DP (70-90% RH)
37°C DP (70-90% RH)
Humidification system output
> 10 mg/l at 10-50 l/min
Compliance
< 1.2 ml/kPa/m
Gas leakage at max. operating pressure
< 10 ml/min
Warm-up time
< 30 min
Environmental conditions
Ambient temperature
10-30°C
Recommended ambient temperature
18°-28°C
Ambient humidity
15-93% RH
Ambient air pressure
700-1060 hPa
Environmental conditions concerning
storage and transport
Temperature
-25 to +70 °C
Humidity
< 93% RH
Air pressure
700-1060 hPa
Electromagnetic compatibility
EN 60601-1-2: 2007
Filter class of dust filter
G4 (EN 779: 2003)
Expected operating time (expected service life)
TNI
soft Flow
50
3-6 years; depends on daily usage
Attached oxygen source
Type
Only medically approved oxygen sources
may be connected (that includes, but is not
limited to oxygen sources complying with
IEC 60601-1:2005). For more information,
please consult the oxygen source user
manual or your oxygen retailer. For handling
and adjustment, please refer to the oxygen
source user manual.
Max. pressure allowed at the oxygen intake
200 mbar