Flowrate too high
Measured flow is higher than the set flow.
Please check air flow.
155
I
Ambient pressure off limits
Ambient pressure is out of permitted range.
Please refer to the product specifications.
156
I
O
2
flow over nominal value
Set O
2
flow rate is too high. Please refer to
chapter 3.1.3 Oxygen.
157
I
Sensor defective
Pressure sensor is defective.*
158
I
Oxygen connection open
Close the oxygen inlet port by a protective
cap or check the proper set-up of the oxygen
source.
201
II
Air flow too hot
Therapy air temperature is too high. Please
check environmental conditions and refer to
the product specifications.
251
I
Ambient temperature off
limits
Ambient temperature is out of permitted ran-
ge. Please refer to the product specifications.
252
I
Ambient humidity off limits
Ambient humidity is out of permitted range.
Please refer to the product specifications.
254
I
Sensor defective
Ambient temperature / humidity sensor is
defective.*
255
I
Dew point not reachable
Set dew point cannot be reached. Please refer
to the product specifications and check the
correct set-up of the system components.
301
II
Heating plate gets too hot
Hardware error.*
302
II
Heating plate defective
Heating plate electronic is not working
properly.*
351
I
Please refill water
Please refill the humidification chamber with
water.
352
I
Heating plate defective
Heating plate is not working properly.*
353
I
Sensor defective
Temperature sensor of humidifier is defecti
-
ve.*
354
I
Heating plate defective
Heating plate is not working properly.*
355
I
Sensor defective
System failure.*
401
II
Blower defective
Blower is blocked.*
402
II
Blower gets too hot
Blower is overheated. Please refer to the pro-
duct specifications and check the air flow.
403
II
Blower sensor defective
Blower temperature sensor is defective.*
404
II
Fan defective
Fan blower is defective.*
501
II
-
Display defective; an acoustic alarm signal is
given.*
502
II
System failure
Sensor errors detected on system startup.*
601
II
Sensor defective
Temperature sensor of applicator is defective;
please replace by a new applicator.