2 Setup
• Use the TNI softFlow system within the given product specifications only.
• If the ambient conditions are out of the required range, keep the device switched off for safety
reasons.
• When the device is brought into the therapeutic environment from outside, a significant tempe-
rature difference (transition from storage conditions to usage conditions) can develop, sometimes
over 50°C. If that has occurred, allow an adaptation to the ambient conditions (room temperature,
e.g.) before startup. This can take 2-4 hours
• Place the TNI softFlow device horizontally on a flat surface below the patient’s head height.
• Place the device at a minimum height of 40 cm from the floor and keep a minimum distance of
40 cm from the wall and 1 m from any other electrical device.
• Place the device so that the power plug can be connected and disconnected without difficulty.
• Use the provided power cable to connect the power socket on the right side of the device to a
power outlet.
• Switch on the device by pressing the rocker switch next to the power socket.
NOTE
• The device performs an internal test during startup: an alarm sound must be audible.
WARNING
• Ensure that the interior of the TNI softFlow unit is dry before connecting it to the power
supply.
2.1 Humidifier
The TNI softFlow system can be operated with two different humidifier types. The TNI softFlow system is
used with the Humidifier Clinic for patients in the clinic or in a care facility, with the Humidifier Homecare
for patients at home. Accordingly, the respective humidifier type must be selected in the user menu, tab
“Humidifier type”.