31
VSI-de-170326
3 Installation
3.1
Notes for Installation
Strong magnetic field!
The gauge contains strong permanent magnets.
Unauthorized modifications or conversions of the instrument are
not allowed!
Installation location
: Indoor
For not fully air conditioned open buildings and operation rooms:
Temperature:
+5°C ... +60°C
Rel. Humidity:
max. 80% up to 30°C, max. 50% at 40°C, not condensing
Air pressure:
860 - 1060 hPa (mbar)
3.2
Vacuum Connection
Dirt and damage, especially at the vacuum flange, have an ad-
verse effect on the function of this vacuum component.
Please take account of the necessary instructions with regard to
cleanliness and damage prevention when using vacuum compo-
nents.
-
Remove the protective cover (is required again during maintenance work!)
-
Make vacuum connection via small flange DN25 ISO KF (VSI17), DN40 ISO
KF (VSI18) or conflat flange DN40CF (VSI19)
-
Use clamps, that can be opened and closed with appropriate tools only (e.g.
strap retainer-tension-ring), use sealing rings with a centering ring
-
Make sure that the sensor flange is connected to ground, e.g. by having elec-
trical contact to the grounded vacuum chamber (use metallic clamps)
The transducer may be mounted in any orientation. Mounting with the flange to
the top, however, can lead to early contamination.
An upright orientation with flange to the bottom is to be preferred in order to keep
particles and condensates out of the sensor cell. Further the transducer is ad-
justed in the upright position ex works.