46
VSI-de-170326
In the software telegram the following parameters will be transmitted:
CHAR:
"Log": logarithmic relation between voltage and pressure
V
out
[V] = GAIN[V] x log (p[mbar]) + OFFSET[V]
"Lin": linear relation between voltage and pressure
V
out
[V] = GAIN[V/mbar] x p[mbar] + OFFSET[V]
GAIN:
Gain in V per pressure decade or V per mbar
OFFSET:
Voltage offset in V
LOW.L:
smallest admissible output voltage in V
UPP.L:
highest admissible output voltage in V
UNDER.:
Voltage output in V in case of pressure underrange
OVER.:
Voltage output in V in case of pressure overrange
FAULT:
Error signal in V
5.8
VacuGraph
TM
Software
VacuGraph software has been especially developed for use with Thyracont
gauges and is available for Windows and Android operating systems.
VacuGraph features plotting and saving of measurement data as well as a com-
fortable way of configuring all device parameters.
Also available as set SLZUB
with protective case, inter-
face adapter RS485-USB
and 24V plug-in power sup-
ply for the transducer!
Download VacuGraph:
www.vacugraph.com
Example features:
-
Plot, analyze and save measurement curves
-
Compare multiple plots
-
Export measurement data for MS Excel
-
Automatic calculation of leak rates by rate-of-rise measurements
-
Easy configuration of all device parameters
-
Scaling wizard with graphic support for adjusting the voltage output
characteristic