43
VSI-de-170326
Setting T[p1] = F[p2] is forbidden!
A gap between the thresholds being too small will result in flick-
ering of the relay and can even cause damage to the relay!
Setting E:
Relay closes in case of a device error.
Setting !E:
Relay opens in case of a device error.
Setting U:
Relay closes in case of pressure underrange.
Setting !U:
Relay opens in case of pressure underrange.
Setting O:
Relay closes in case of pressure overrange.
Setting !O:
Relay opens in case of pressure overrange.
Setting C:
Relay closes when cold cathode is switched on.
Setting !C:
Relay opens when cold cathode is switched on.
Setting T0:
Open relay by software command (off).
Setting T1:
Close relay by software command (on).
The control LEDs on top of the transducer will be on as soon
as the related relay is closed.