4 General Information
_________________________________________________________________________________
______________________________________________________________________________________________________
© 2012 Thorlabs
Attention
Mobile telephones, cellular phones or other radio transmitters are not to be
used within the range of three meters of this unit since the electromagnetic
field intensity may then exceed the maximum allowed disturbance values
according to EN 50 082-1.
This equipment has been tested and found to comply with the limits for a Class
A digital device, pursuant to part 15 of the FCC Rules. These limits are
designed to provide reasonable protection against harmful interference when
the equipment is operated in a commercial environment. This equipment
generates, uses, and can radiate radio frequency energy and, if not installed
and used in accordance with the instruction manual, may cause harmful
interference to radio communications. Operation of this equipment in a
residential area is likely to cause harmful interference in which case the user
will be required to correct the interference at his own expense.
Mounting several PRO8000 (-4) in a rack, make sure to have 1 height-unit space
between the different housings to ensure proper ventilation (ventilation slots
are beneath the front panel!!)
To connect the unit to the mains, use only the supplied mains cord with safety mains
plug. The unit is grounded via the protective conductor of this cable. To avoid electric
shocks the plug of the mains cable must be inserted in a correctly grounded socket.
Interruption of the protective grounding could lead to health damage or even death
due to electric shock.
Please check prior to operation, if the line voltage, indicated on the rear panel agrees
with your local supply. If not please have a service technician change the line voltage
and the appropriate fuse.
Attention
Changing the line voltage is a service operation and must be done only by
qualified service personnel! (see section 6, Service).
To guarantee safe operation of the PRO8000 (-4) / PRO800 do not obstruct the
ventilation slots beneath the front panel and the fan air outlets at the rear panel.