62 Communication with a PC
_________________________________________________________________________________
______________________________________________________________________________________________________
© 2012 Thorlabs
Errors while using the data read-out commands “
:ELCH:GETALL?
” or
“
:ELCH:TRIG?
”
e.g.: “
Timeout expired before operation completed
”
or
e.g.: “
The number of bytes transferred is equal to
the input count
”
Timeout value to small .
Input buffer too small.
You don’t get out all measured data : Perhaps your measurement is still
in progress. Either test the ELCH status (see 4.7.7) or use the
“
:ELCH:GETALL?
” command again to retrieve the rest of the data.
Your retrieved data do not start with the expected first set value:
Maybe you had another ELCH run previous and did not read out the old
data.
Start your ELCH run with the
:ELCH:RESET 0
command, to discard
previous measured values.