70 Maintenance and repair
_________________________________________________________________________________
______________________________________________________________________________________________________
© 2012 Thorlabs
The PRO8 enters remote mode with the first RS232 command, but does not respond
further
Do you have selected the corresponding baud rate?
Either change the baud rate in the PRO8 setup menu or in your application
program.
The key “LOCAL” on the PRO8000/800 keypad is not working
Your remote control computer may have send a “LLO” (local lockout) message. The
“LOCAL” key is then deactivated. To enter local mode again, your PC must send a
“GTL” (go to local) message. The deactivation of the “LOCAL” key remains until you
turn off your PRO8000 (-4) / PRO800 system.
Displayed measured values seem to react only very slowly
Adjust the over sampling rate (OSR) in the setup menu to a lower value.
If you don’t find the error source by means of the trouble shooting list or if more modules
work erratic please contact
Thorlabs
for technical support and/or return instructions.