16 Operating the PRO8000 (-4) / PRO800
_________________________________________________________________________________
______________________________________________________________________________________________________
© 2012 Thorlabs
3.5 Setup menu
From main screen you can enter the set-up screen by pressing
.
PRO8000 SETUP
Beeper : on
Interface:IEEE488
IEEE Adr : 10
Baudrate : 19200
OSR local
1024
CHANGE
The first line marked and hitting “CHANGE” toggles the beeper for occurring errors
and input confirmation on and off.
In the second line you can select between IEEE488 Interface or terminal mode via
RS232 Interface.
In the third line the IEEE-address is set, default value is 10.
The appropriate baud rate for the RS 232-interface is chosen in line 4 (default 19200
Baud).
OSR local
allows to select the averaging rate for displayed values, i.e. how many
measured values are averaged to reduce noise. The higher the value, the longer it
takes to actualize the measured values on the screen. This item does not affect the
remotely transferred measurement values. (‘Normal’ value =16).
. Brings you back to main menu.