![THORLABS PRO800 Series Operation Manual Download Page 70](http://html.mh-extra.com/html/thorlabs/pro800-series/pro800-series_operation-manual_1110711070.webp)
66 Maintenance and repair
_________________________________________________________________________________
______________________________________________________________________________________________________
© 2012 Thorlabs
Addresses”, on page 85.
5.6 Installing and uninstalling modules
Attention
Modules are only to be installed or uninstalled with the mainframe
switched off. If not, severe damage could occur to module and/or main-
frame!!
The PRO8000 (-4) / PRO800 has a modular design and can be operated with a variety of
different modules.
The change of modules or retro-fitting of modules can be done by the user himself.
It is not necessary to re-adjust the modules after reconfiguration of the PRO8000 (-4) /
PRO800.
The number of modules that can be used simultaneously is limited by the number of slots.
The modules can be freely configured in all slots.
After switching on the PRO8000 (-4) / PRO800 will recognize each new configuration
automatically.
Uninstalling a module
Switch the PRO8000 (-4) / PRO800 off and remove the mains cable from the mains
jack. The modules are uninstalled from the front of the unit.
Modules of normal width are fixed with two screws, modules with double or triple width
with four screws.
Pull out the modules by means of the handle.
Close the free slot with a blind plate, or install another module (see below).
Check the function of the PRO8000 (-4) / PRO800 by switching it on.
Installing a module
Depending on the required width of the module a sufficient number of slots must be
free; uninstall modules (see above) or remove blind plates.
Put the module into the guide rail of the slot and push until it locks in.
Fix the modules with the appropriate number of screws.
Check the function of the modules by switching the PRO8000 (-4) / PRO800 on.
The inserted modules must be recognized correctly in the set-up menu.