3.7 Error messages of the ITC5xx
ITC500 / page 88
[
1308, "No setting of IMD during constant current
mode"
]
Category:
Execution
Error
Possible reason:
The set value of the monitor diode current and the optical power
cannot be changed during constant current mode.
[
1309, "No LD polarity change during laser on"
]
Category:
Execution
Error
Possible reason: The polarity of the laser diode can not be changed during the
laser diode output switched on.
[
1310, "No PD polarity change during laser on"
]
Category:
Execution
Error
Possible reason:
The polarity of the monitor diode can not be changed during the
laser diode output switched on.
[
1311, "No mode change during laser on"
]
Category:
Execution
Error
Possible reason: The operating mode CC or CP cannot be changed during the
laser diode output is switched on.
[
1312, "Wrong or no sensor"
]
Category:
Execution
Error
Possible reason:
The ITC5xx detected no sensor at the output jack.
[
1313, "Wrong command for this sensor"
]
Category:
Execution
Error
Possible reason:
The released command does not fit to the connected sensor.
[
1314, "No sensor change during TEC on allowed"
]
Category:
Execution
Error
Possible reason: The type of sensor may not be changed during the output
switched on.