3.6 ITC5xx specific commands
ITC500 / page 74
3.6.6 Turn the output on and off (LASER)
Programming:
Syntax:
":LASER OFF"
":LASER ON"
Default Value: OFF
Assumption: To switch the output on there must be no device errors (interlock
open, open circuit, over temperature, ...).
Description: Switches the laser diode output on or off.
Reading:
Syntax:
":LASER?"
Response:
[:LASER OFF]
[:LASER ON]
Description: Queries the state of the laser diode output.
NOTE
To switch on the laser while a “open circuit“ error is present (see Device
Error Condition Register), you have to reset the “open circuit“ error with
the command:
“:LASER ON“
. With the next command
“:LASER ON“
the laser will switch on again.