3.7 Error messages of the ITC5xx
ITC500 / page 87
3.7.2 ITC5xx operation error messages
[
1301, "Interlock is open"
]
Category:
Execution
Error
Possible reason:
Try to switch on the output while the Interlock line is open.
(Refer to chapter 2.22, "Using the interlock input" on page 52)
[
1302, "Open circuit"
]
Category:
Execution
Error
Possible reason:
Try to switch on the output while the able to the laser diode has
opened or the compliance voltage is not high enough.
[
1303, "Over temperature"
]
Category:
Execution
Error
Possible reason: Try to switch on the output while the internal temperature is too
high. Wait until the ITC5xx has cooled down. Maintain proper air
flow.
[
1304, "Internal power failure"
]
Category:
Execution
Error
Possible reason: Try to switch on the output while an internal power failure
occurred.
[
1306, "No calibrating of PD during laser on"
]
Category:
Execution
Error
Possible reason: The monitor diode cannot be calibrated during the laser diode
output is switched on.
[
1307, "No setting of ILD during constant power mode"
]
Category:
Execution
Error
Possible reason: The set value of the laser diode current can not be changed
during constant power mode.