![THORLABS ITC502 Operation Manual Download Page 20](http://html.mh-extra.com/html/thorlabs/itc502/itc502_operation-manual_1110749020.webp)
1.5 Technical Data
ITC500 / page 12
Analog modulation input
Input impedance
10 k
Ω
Small signal 3 dB bandwidth (CC mode)
DC ... 200 kHz
Laser diode modulation coefficient (CC mode)
100 mA/V ± 5%
Laser diode modulation coefficient (CP mode)
0.2 mA/V ± 5%
Current output TEC element
Control range
- 4 A ... + 4 A
Maximum output power
32 W
Compliance voltage
>
8 V
Measurement resolution (manual / remote control)
1 mA / 0.1 mA
Measurement accuracy
± 20 mA
Noise and ripple (typ.)
<
2 mA
Measurement resolution TEC voltage (manual / remote control)
1 mV / 0.1 mV
Measurement accuracy TEC voltage
± 40 mV
TEC Current limit
Setting range
0 ...
≥
4 A
Measurement resolution (manual / remote control)
1 mA / 0.1 mA
Accuracy
±
40 mA
Temperature sensors AD590, AD592, LM135, LM335
Control range (AD590, LM135)
- 45 °C ... + 145 °C
Control range (AD592)
- 25 °C ... + 105 °C
Control range (LM335)
- 40 °C ... + 100 °C
Setting resolution (manual / remote control)
0.01 °C / 0.003 °C
Measurement resolution (manual / remote control)
0.01 °C / 0.001 °C
Accuracy
±
0.1 °C
Temperature stability (24 hours)
≤
0.001 °C
Summary of Contents for ITC502
Page 2: ...Version 2 16 Date 25 07 2006 Copyright 2006 Thorlabs ...
Page 8: ......