![THORLABS ITC502 Operation Manual Download Page 91](http://html.mh-extra.com/html/thorlabs/itc502/itc502_operation-manual_1110749091.webp)
3.6 ITC5xx specific commands
ITC500 / page 83
3.6.19 Reading the TEC current hardware limit (LIMTP)
Syntax:
":LIMTP:ACT?"
Response:
[:LIMTP:ACT <NR3>]
Description: Queries the actual TEC current hardware-limit. Unit: [A]
(Refer to Chapter 1.8.2, "Setting the limit value of the TEC current" on page 16)
3.6.20 Reading the TEC current (ITE)
Syntax:
":ITE:ACT?"
Response:
[:ITE:ACT <NR3>]
Description: Queries the actual TEC current. Unit: [A]
3.6.21 Reading the temperature/resistance window (WIN)
Syntax:
":WIN:ACT?"
Response:
[:WIN:ACT <NR3>]
Description: Queries the actual temperature/resistance window. Depending on the
temperature sensor the temperature window or the resistance window
is retuned. Unit: [°C], [
Ω
]
3.6.22 Reading the TEC voltage (VTE)
Syntax:
":VTE:ACT?"
Response:
[:VTE:ACT <NR3>]
Description: Queries the actual TEC voltage. Unit: [V]
Summary of Contents for ITC502
Page 2: ...Version 2 16 Date 25 07 2006 Copyright 2006 Thorlabs ...
Page 8: ......