1.8 Pre-settings
ITC500 / page 18
SW4
defines if the bias voltage is automatically switched on:
SW4
function
up
bias off
down
bias on
Attention
Before activating this function make sure, that the monitor diode
polarity is set correctly when switching on
(SW4 must be set correctly).
If the monitor diode is forward biased with a voltage the resulting
current may damage or even destroy the monitor diode (= the laser
module).
SW5
defines whether the laser diode temperature window protection is on or off,
indicated by the LED WIN in the field
SETUP
:
SW5
function
up
WIN off
down
WIN on
SW6
defines which potentiometer for the TEC current is active:
SW6
function
up
VAR