![THORLABS ITC502 Operation Manual Download Page 77](http://html.mh-extra.com/html/thorlabs/itc502/itc502_operation-manual_1110749077.webp)
3.4 System command group
ITC500 / page 69
3.4.3 Querying the error queue
Syntax:
":SYST:ERR?"
Response:
[0, "No Error"]
Description: Queries the error queue of the ITC5xx. The reply consists of the
following sequence: <Error No.>, "<Error text>". If the error queue is
empty:
[0, "No error"]
is given.
(Please refer to chapter 3.7, “Error messages of the ITC5xx“, starting on page 84)
3.4.4 Averaging rate
Programming:
Syntax:
":SYST:OSR <NR1>"
Valid Range: 0..7
Default Value: 5
Description: Sets the ITC5xx averaging rate (see table below).
Reading:
Syntax:
":SYST:OSR?"
Response:
[<NR1>]
Description: Queries the averaging rate.
Sampling rate
Resolution
Measurement speed
0
11 Bit + sign
18 / s
1
12 Bit + sign
13 / s
2
13 Bit + sign
10 / s
3
14 Bit + sign
8 / s
4
15 Bit + sign
6 / s
5
16 Bit + sign
4 / s
6
17 Bit + sign
3 / s
7
18 Bit + sign
2 / s
Summary of Contents for ITC502