Alarms and Troubleshooting
7
HeartMate III Left Ventricular Assist System Instructions for Use
7-9
Table 7.1 System Controller Advisory Alarms (Continued)
IMPORTANT!
The Pump Running (
) light is always green on when the Pump is running.
Priority
System
Controller
Screen
Active
Symbols
Silence
Period
Alarm Means
To Resolve Alarm
ADVISOR
Y
4 hours
System
Controller
Backup
Battery not
installed
1. Install the 11 Volt Lithium-Ion
backup battery in the System
Controller.
2. Obtain a new backup battery
replacement kit.
Note:
If replacing the battery does
not resolve the alarm, the System
Controller may need to be replaced,
or additional steps may be required.
Call Thoratec Corporation with
questions. Refer to page 7-22.
4 hours
Driveline Power
Fault
Contact Thoratec Corporation to
determine best next steps. Use the
System Monitor to silence the alarm
while awaiting resolution, if needed.
Note:
The alarm must be active in
order to access the alarm silence for
this situation. Refer to page 7-23.
4 hours
Driveline
Communication
Fault (Driveline
Comm Fault)
Contact Thoratec Corporation to
determine best next steps.
Use the System Monitor to silence
the alarm while awaiting
resolution, if needed.
Note:
The alarm must be active to
access the alarm silence for this
situation. Refer to page 7-24.
+
Summary of Contents for HeartMate III
Page 1: ...Instructions for Use ...
Page 2: ......
Page 3: ...Thoratec Corporation HEARTMATE III LEFT VENTRICULAR ASSIST SYSTEM Instructions for Use ...
Page 4: ...HEARTMATE III LEFT VENTRICULAR ASSIST SYSTEM INSTRUCTIONS FOR USE ii ...
Page 6: ...HEARTMATE III LEFT VENTRICULAR ASSIST SYSTEM INSTRUCTIONS FOR USE iv ...
Page 10: ...Contents viii HEARTMATE III LEFT VENTRICULAR ASSIST SYSTEM INSTRUCTIONS FOR USE ...
Page 12: ...1 Introduction 1 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 28: ...1 Introduction 1 18 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 30: ...2 System Operations 2 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 88: ...3 Powering the System 3 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 112: ...3 Powering the System 3 26 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 166: ...4 System Monitor 4 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 208: ...4 System Monitor 4 44 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 210: ...5 Surgical Procedures 5 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 378: ...A A 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 390: ...B B 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 420: ...C C 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 426: ...C C 8 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 428: ...D D 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 448: ...E E 10 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 457: ......