3
Powering the System
3-36
HeartMate III Left Ventricular Assist System Instructions for Use
• The Mobile Power Unit has an AC Power Cord and Patient Cable, both of which may be a
tripping hazard. Ensure that the patient, caregivers, and all other persons near the Mobile
Power Unit are aware of this potential hazard.
• Do not clean or service the Mobile Power Unit while it is plugged into an AC electrical
outlet, or electrical shock may occur.
• Do not incinerate, disassemble, crush, puncture, or otherwise damage batteries, as this
can cause leakage or rupture, resulting in personal injury or damage to the Mobile Power
Unit.
• Do not mix old and new batteries, or mix battery types (such as rechargeable and
non-rechargeable). This can cause battery leakage or rupture, resulting in personal injury
or damage to the Mobile Power Unit.
• Mobile Power Unit power output may be affected by mobile phones, resulting in low
power alarms on the System Controller, or loss of the green power LED on the Mobile
Power Unit. If either of these conditions is observed, separate the mobile phone from the
Mobile Power Unit by at least .6 meters (24 inches). If the condition persists after
separating the devices, switch to two HeartMate 14 Volt Lithium-Ion batteries.
• Keep the Mobile Power Unit free of excessive lint and dust. Also, keep the Mobile Power
Unit away from heat or humidity sources, such as a fireplace, radiant heater, nebulizer, or
steam kettle. The Mobile Power Unit may fail to operate properly due to excessive lint,
dust, heat, or humidity.
• Inspect the Mobile Power Unit patient and power cables for damage. Do not use the
Mobile Power Unit if either cable shows signs of damage.
• When moving the Mobile Power Unit to a different location or AC power source, first
connect the System Controller to HeartMate 14 Volt batteries.
• Do not change the Mobile Power Unit batteries while the Mobile Power Unit is powering
the HeartMate system. Prior to replacing the Mobile Power Unit batteries, switch to another
power source. Then disconnect the Mobile Power Unit power cord from the wall socket.
• Do not carry or touch the Mobile Power Unit for an extended time. To avoid the risk of
burns, do not touch the top surface of the Mobile Power Unit for longer than one minute.
The Mobile Power Unit surface temperature can become uncomfortably warm, especially
when the room temperature is above 104°F (40°C). Surface temperatures can approach
131°F (55°C).
CAUTION !
Summary of Contents for HeartMate III
Page 1: ...Instructions for Use ...
Page 2: ......
Page 3: ...Thoratec Corporation HEARTMATE III LEFT VENTRICULAR ASSIST SYSTEM Instructions for Use ...
Page 4: ...HEARTMATE III LEFT VENTRICULAR ASSIST SYSTEM INSTRUCTIONS FOR USE ii ...
Page 6: ...HEARTMATE III LEFT VENTRICULAR ASSIST SYSTEM INSTRUCTIONS FOR USE iv ...
Page 10: ...Contents viii HEARTMATE III LEFT VENTRICULAR ASSIST SYSTEM INSTRUCTIONS FOR USE ...
Page 12: ...1 Introduction 1 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 28: ...1 Introduction 1 18 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 30: ...2 System Operations 2 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 88: ...3 Powering the System 3 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 112: ...3 Powering the System 3 26 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 166: ...4 System Monitor 4 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 208: ...4 System Monitor 4 44 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 210: ...5 Surgical Procedures 5 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 378: ...A A 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 390: ...B B 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 420: ...C C 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 426: ...C C 8 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 428: ...D D 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 448: ...E E 10 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 457: ......