Powering the System
3
HeartMate III Left Ventricular Assist System Instructions for Use
3-65
12. Complete the following steps:
a. Connect the black Mobile Power Unit patient cable connector to the black System
Controller power cable connector.
b. Tighten the nut.
The alarm will stop.
13. Press the battery release button on one of the battery clips to release its battery.
14. Repeat Step 13 for the second battery.
The System Controller is now connected to the Mobile Power Unit, and the Mobile Power
Unit is powering the system.
15. Store the battery clips in a clean, dry location until the next use.
16. Place the depleted batteries into the Battery Charger for charging.
Refer to
Battery Charger Overview
The Battery Charger (
) is designed to charge the HeartMate 14 Volt Lithium-Ion
batteries that are used to power the HeartMate III Left Ventricular Assist System during
battery-powered operation. Specifically, the Battery Charger can:
• Charge up to four batteries in four hours or less. Refer to
• Monitor the need for calibration, and calibrate individual batteries. Refer to
• Perform diagnostic testing on up to four batteries at a time. Refer to
Figure 3.70 Battery Charger
Summary of Contents for HeartMate III
Page 1: ...Instructions for Use ...
Page 2: ......
Page 3: ...Thoratec Corporation HEARTMATE III LEFT VENTRICULAR ASSIST SYSTEM Instructions for Use ...
Page 4: ...HEARTMATE III LEFT VENTRICULAR ASSIST SYSTEM INSTRUCTIONS FOR USE ii ...
Page 6: ...HEARTMATE III LEFT VENTRICULAR ASSIST SYSTEM INSTRUCTIONS FOR USE iv ...
Page 10: ...Contents viii HEARTMATE III LEFT VENTRICULAR ASSIST SYSTEM INSTRUCTIONS FOR USE ...
Page 12: ...1 Introduction 1 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 28: ...1 Introduction 1 18 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 30: ...2 System Operations 2 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 88: ...3 Powering the System 3 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 112: ...3 Powering the System 3 26 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 166: ...4 System Monitor 4 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 208: ...4 System Monitor 4 44 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 210: ...5 Surgical Procedures 5 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 378: ...A A 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 390: ...B B 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 420: ...C C 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 426: ...C C 8 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 428: ...D D 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 448: ...E E 10 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 457: ......