Powering the System
3
HeartMate III Left Ventricular Assist System Instructions for Use
3-61
Switching from Battery Power to the Power Module
Use care when connecting or disconnecting power cables. For more information, refer to
Guidelines for Power Cable Connectors
F
OR
THIS
TASK
YOU
NEED
:
• 1 running System Controller on 14 Volt Lithium-Ion battery power
• 1 functioning Power Module
• Functioning and grounded (3-prong) AC electrical outlet that is dedicated to Power Module
use and not controlled by a wall switch
• 1 AC power cord to connect to an AC electrical power outlet
• 1 battery holster or other wear and carry accessory
T
O
SWITCH
FROM
B
ATTERIES
TO
THE
P
OWER
M
ODULE
:
1. Confirm that the Power Module is plugged into a properly tested and grounded (3-prong) AC
electrical outlet that is dedicated to Power Module use.
2. Perform a Power Module self test and complete one of the following steps:
• If the Power Module passes the self test, go to Step 3
• If the Power Module fails the self test, contact Thoratec Corporation
Refer to
Performing a Power Module Self Test
3. Complete one of the following steps:
• If the Patient Cable is connected to the Power Module, continue to Step 5.
• If the Patient Cable is not connected to the Power Module, go to Step 4.
4. Complete the following steps to connect the Patient Cable to the Power Module:
a. Line up the red dot on the Patient Cable with the red dot near the "
" socket on the Power
Module.
b. Insert the Power Module patient cable into the socket (
WARNING !
• Do not use an outlet that is controlled by a wall switch.
• Do not use an adapter plug for an ungrounded wall outlet.
• Do not use portable, multiple outlet (power strip) adapters.
• Using adapters or power strips may cause a serious electrical shock or cause the Pump to
stop.
Summary of Contents for HeartMate III
Page 1: ...Instructions for Use ...
Page 2: ......
Page 3: ...Thoratec Corporation HEARTMATE III LEFT VENTRICULAR ASSIST SYSTEM Instructions for Use ...
Page 4: ...HEARTMATE III LEFT VENTRICULAR ASSIST SYSTEM INSTRUCTIONS FOR USE ii ...
Page 6: ...HEARTMATE III LEFT VENTRICULAR ASSIST SYSTEM INSTRUCTIONS FOR USE iv ...
Page 10: ...Contents viii HEARTMATE III LEFT VENTRICULAR ASSIST SYSTEM INSTRUCTIONS FOR USE ...
Page 12: ...1 Introduction 1 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 28: ...1 Introduction 1 18 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 30: ...2 System Operations 2 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 88: ...3 Powering the System 3 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 112: ...3 Powering the System 3 26 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 166: ...4 System Monitor 4 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 208: ...4 System Monitor 4 44 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 210: ...5 Surgical Procedures 5 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 378: ...A A 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 390: ...B B 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 420: ...C C 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 426: ...C C 8 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 428: ...D D 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 448: ...E E 10 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 457: ......