4
System Monitor
4-16
HeartMate III Left Ventricular Assist System Instructions for Use
Advisory Alarms
Advisory alarms are used to inform users about conditions that affect optimal system operation.
Although important, Advisory alarms are not related to life threatening risks. Advisory alarms
appear on the System Monitor screen as black text on a yellow banner. They are also
accompanied by audible alarms on the System Controller.
There are eleven Advisory alarms. The on-screen appearances of each Advisory alarm is
described as follows, in order of descending priority:
• Power Cable Disconnect: This text banner is accompanied by one beep every second.
• Low Voltage Advisory: This text banner is accompanied by one beep every four seconds.
• Replace System Controller: A text banner appears; and in most cases is accompanied by one
beep every four seconds. There may be instances when there is no audible alarm.
• Communication Fault (Comm Fault): This text banner is accompanied by one beep every four
seconds.
• Replace Backup Battery: A text banner appears; and in most cases is accompanied by one
beep every four seconds. There may be instances when there is no audible alarm.
• WARNING: Low Speed Advisory: A text banner appears; however, there is no audible alarm.
• Backup Battery Not Installed: This text banner is accompanied by one beep every four seconds.
• LVAD Fault: A text banner appears; however, there is no audible alarm.
• Driveline Power Fault: This text banner is accompanied by one beep every four seconds.
• Driveline Communication Fault (Driveline Comm Fault): This text banner is accompanied by one
beep every four seconds.
• Controller Clock Not Set: A text banner appears; however, there is no audible alarm.
During alarm conditions, a
Silence Alarm
button appears in the lower right corner of the
screen. Press this button to temporarily silence audible alarms. Refer to
• The Silence Alarm button may still appear for alarm conditions for which there is no associated
audible alarm. In this case, the button will not perform any function.
• Hazard alarms and the Power Cable Disconnect alarm can be silenced for two minutes.
• The Low Voltage Advisory can be silenced for 5 minutes.
• All other Advisory alarms with audible tones can be silenced for four hours. The exceptions are
Communication Fault, Driveline Power Fault, Driveline Communication Fault, which can be
silenced indefinitely or for 24 hours.
Summary of Contents for HeartMate III
Page 1: ...Instructions for Use ...
Page 2: ......
Page 3: ...Thoratec Corporation HEARTMATE III LEFT VENTRICULAR ASSIST SYSTEM Instructions for Use ...
Page 4: ...HEARTMATE III LEFT VENTRICULAR ASSIST SYSTEM INSTRUCTIONS FOR USE ii ...
Page 6: ...HEARTMATE III LEFT VENTRICULAR ASSIST SYSTEM INSTRUCTIONS FOR USE iv ...
Page 10: ...Contents viii HEARTMATE III LEFT VENTRICULAR ASSIST SYSTEM INSTRUCTIONS FOR USE ...
Page 12: ...1 Introduction 1 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 28: ...1 Introduction 1 18 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 30: ...2 System Operations 2 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 88: ...3 Powering the System 3 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 112: ...3 Powering the System 3 26 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 166: ...4 System Monitor 4 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 208: ...4 System Monitor 4 44 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 210: ...5 Surgical Procedures 5 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 378: ...A A 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 390: ...B B 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 420: ...C C 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 426: ...C C 8 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 428: ...D D 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 448: ...E E 10 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 457: ......