11.2.11
Command KY
<id>KY<parameter><CR>
Key/password
Access:
reading / writing
Description:
With this command, the value for the key (password) is set. The
following 3 password levels are possible:
Query (reading)
User (general settings)
Calibration data (access to the calibration data sets)
KX (general settings / access to the calibration data sets)
Parameter type:
unsigned integer
Parameter:
Parameter
Description
0
Query
234
1
User
137
1
Calibration data (see command CK)
KX
2
User defined password
1)
The key „234“ and „137“ are valid, if parameter KX the value 0 in-
clude.
2)
Include parameter KX the value unlike 0, then replace the value both
keys „234“ (User) and „137“ (Calibration data).
Type return value:
unsigned integer
Return value:
see parameter
Value range:
0, 137, 234, KX
Initial value:
0
Note:
After changing the parameter "KX" to a value other than 0, all parameters
(general settings and calibration data) can only be changed by specifying
the password set with "KX"!
Note:
Please contact the manufacturer if you have lost the password ("KX")!