11.2.3 Command CI
<id>CI<parameter><CR>
Selection of the command interpreter.
Access:
reading / writing
Description:
With the command CI, the desired command interpreter is set.
Note:
If the identification number (ID) is bigger than 98, it is au-
tomatically set to 0 when switching over into the THIES in-
terpreter!
Note:
If the identification number (ID) equals 0, a switch-over
into the MODBUS-RTU interpreter is not possible!
Parameter description:
Parameter
Description
0
THIES
1
MODBUS RTU
Value range:
0 to 1
Initial value:
0/1 depending on device version
0 for 4.3352.x0.400, 1 for 4.3352.x0.401
11.2.4 Command CK
<id>CK<parameter><CR>
Password for writing access to calibration tables.
Access:
writing
Description:
With the command, the password is set for writing access to the
calibration tables (see command WC). The password must be
set 2 consecutive times within a time frame of 60 seconds.
If the password has been changed once, the current password
must be set before changing it again.
Parameter type:
unsigned integer
Parameter:
1…4294967295
Type return value:
unsigned integer
Return value:
see parameter
Value range:
1…4294967295
Initial value:
137