11.2.10
Command KX
<id>KX<parameter><CR>
With the command KX the command KY password can be
changed
Access:
writing
Description:
The command KX sets the password for write access to the pa-
rameters of command KY. A value other than 0 overwrites or
replaces the password levels "User" and "Calibration data" (see
command KY).
The password must be set 2 times in succession, in a time win-
dow of 60 seconds.
Once the password has been changed, the current password
must be set before changing again. The password is entered
using the KY command. If the parameter for the KX command
is 0, then the password levels “User” and “Calibration data”
listed for the KY command are valid.
Parameter type:
unsigned integer
Parameter:
0…4294967295
Type return value:
unsigned integer
Return value:
see parameters
Value range:
0…4294967295
Initial value:
0 (please note following note)
Sample:
1) Key „User“ set
00KY234
! 00KY234
2) User defined password set to 1“(1. Set)
00KX1
3) User defined password set to 1 (2. Set)
00KX1
!00KX00001
4) Test the user defined password to chance the ID
00KY1
!00KY00001
00ID1
!01ID00001
Note:
After updating the firmware to version V3.02, the parameter "KX" contains
the value 4294967295!