Operation
Settings
3-76
43iQHL Instruction Manual
Thermo Scientific
●
Lamp Intensity:
Displays the current lamp intensity.
●
Down:
●
Manual Entry:
Shows current value. If pressed, user can manually
set the detector gain value.
●
Auto Cal:
When pressed, automatically calibrates the Detector. The
detector’s auto-calibration process is initiated. Please allow up to 5
minutes for calibration to complete. User can stop calibration by
pressing the Stop button under Detector Gain.
●
End Cal:
When pressed, the auto calibration is interrupted and the
value does not change.
●
Default Values:
When pressed, the default gain for the lamp
intensity detector is restored.
Home Screen>Settings>Measurement Settings>Advanced Measurement
Settings>Optical Bench Settings>Continue To PMT Settings Screen
The PMT Settings screen contains the following information:
●
Across:
●
Description:
Defines the actions the user can do.
●
PMT Supply:
User can manually adjust the PMT supply value, read
target and calibration concentration, start/stop automatic
calibration, read voltage, frequency and zero offset frequency
values, and reset default values.
●
Current PMT:
Displays the current PMT voltage value.
●
Units:
Displays units for each item.