Operation
Settings
3-42
43iQHL Instruction Manual
Thermo Scientific
The SO
2
Bench screen provides status and alarms related to the SO
2
bench
module. If an item being monitored goes outside the lower or higher alarm
limit, an alarm is activated.
Use the
and
buttons to move up and down and the
and
buttons to move left and right.
Home Screen>Settings>Health Check>Status and Alarms>SO
2
Bench
The SO
2
Bench screen contains the following information:
●
Across:
●
SO
2
Bench:
This column lists items associated with the SO
2
bench.
●
Value:
Displays the current value for each item.
●
Low Alarm:
Displays low alarm for each item.
●
High Alarm:
Displays high alarm for each item.
●
Units:
Displays units for each item.
●
Down:
●
Chamber Temperature:
Displays the current chamber temperature.
User can adjust low and high alarm limits.
●
Lamp Intensity:
Displays the current lamp intensity reading. User
can adjust low and high alarm limits.
●
Lamp Voltage:
Displays the current lamp voltage. User can adjust
low and high alarm limits.
●
PMT Voltage:
Displays the current PMT supply voltage.
●
PMT Voltage Too Low:
Displays OK/Fail for PMT voltage too low.
●
PMT Voltage Too High:
Displays OK/Fail for PMT voltage too
high.
SO
2
Bench
Summary of Contents for 43iQHL
Page 1: ...43iQHL Instruction Manual High Level Sulfur Dioxide Analyzer 119015 00 1Apr2022...
Page 2: ......
Page 8: ......
Page 18: ......
Page 34: ...Operation Instrument Display 3 8 43iQHL Instruction Manual Thermo Scientific...
Page 150: ......
Page 176: ......
Page 238: ......
Page 256: ......
Page 264: ......
Page 276: ......
Page 277: ......