Operation
Settings
3-74
43iQHL Instruction Manual
Thermo Scientific
The Advanced Measurement Settings screen allows the user to calibrate the
optical bench and set other advanced settings.
Home Screen>Settings>Measurement Settings>Advanced Measurement
Settings
The Advanced Measurements Settings screen contains the following
information:
●
Optical Bench Settings:
Calibrates detector gain, PMT supply, flash
voltage, and initial flash reference.
●
Flash Lamp:
Enables/Disables flash lamp.
●
Optical Span Test:
Enables/Disables test LED and reads the LED SO
2
concentration.
●
Extended Ranges:
Enables/Disables extended ranges feature. When
changing to extended range from standard range, the PMP voltage
must be readjusted. For more information on readjusting the PMT
voltage, go to Settings>Measurement Settings>Advanced Measurement
Settings>Optical Bench Settings>
Continue to PMT Settings
.
●
Compensation:
Allows the user to compensate for changes in
temperature and pressure concentration.
●
Pressure Calibration:
Calibrates pressure.
Advanced Measurement
Settings