Optional Equipment
Connecting External Devices
Thermo Scientific
43iQHL Instruction Manual
9-11
The digital I/O board consists of:
●
16 Digital Inputs (18 pin connector)
●
10 Digital Relay Switches (20 pin connector)
●
8 Valve Driver Outputs (16 pin connector)
The digital inputs are TTL (3 V or 5 V) compatible and are pulled high
within the instrument. The active state can be user defined in firmware.
●
Logic Low Threshold: 0.8 V
●
Logic High Threshold: 2.0 V
●
Absolute allowable input voltages: -0.5 to 5.5 V
Figure 9–6.
Digital Inputs
Digital I/O Board
Digital Inputs
Summary of Contents for 43iQHL
Page 1: ...43iQHL Instruction Manual High Level Sulfur Dioxide Analyzer 119015 00 1Apr2022...
Page 2: ......
Page 8: ......
Page 18: ......
Page 34: ...Operation Instrument Display 3 8 43iQHL Instruction Manual Thermo Scientific...
Page 150: ......
Page 176: ......
Page 238: ......
Page 256: ......
Page 264: ......
Page 276: ......
Page 277: ......