Internal printer
Installing the internal printer
|
|
|
WARNING
|
|
|
|
Turn off the instrument and disconnect the
|
|
power cord from the outlet before installing the
|
|
internal printer.
|
1.
Loosen the captive screw (#1, Figure 6) on the
lamp door by rotating it counterclockwise about
¼ turn.
2.
Open the lamp door.
3.
Use a pen or screwdriver to lift the tabs holding
the door to the hinge (#2 & #3, Figure 6).
4.
Slide the door off the hinge.
5.
Remove the printer (already installed on the
printer door) from its packing.
6.
Unclip the connector wires on the door hinge.
7.
Lower the hinge so it is out of the way.
8.
Connect the wires (#4, Figure 6) and press into
place with a small screwdriver. There is only
one way that the connectors will fit. Each
connector has a slight D shape. Make sure the
side of the connector with the shiny metal
contacts (see Figure 6) and faces away from
the printer and towards the plastic door.
9.
Use the clip on the hinge to secure the wires.
10.
Install the printer door by sliding it back onto
the hinge (#3, Figure 6).
11.
Close the lamp door.
12.
Tighten the captive screw (#1, Figure 6) on the
printer door to hold it securely in place.
13.
Load the paper into the printer (
see instructions
in Setting up the instrument
).
Loading the paper in the internal printer
For instructions on loading the paper in the internal
printer, see
Setting up the instrument
.
External computers
To connect an external computer, make sure that
the pin-outs on the connectors are set according to
the illustration below.
To connect an external computer, connect the
computer cable to the 9-pin RS232C port on the
back of the instrument (#2, Figure 2). Be sure to
use a DB9 pin (female) to DB9 pin (female)
interface cable (335942).
Contact technical support for more information
about using an external computer with the
spectrophotometer.
5-3
BioMate 3 Operator’s Manual
Connecting & Using Accessories
Protocol to use with external device:
Baud rate
19,200
Data bits:
8
Parity:
Off
Stop Bit:
1
Summary of Contents for 335904
Page 1: ...BioMate 3 Series Spectrophotometers Operator s Manual ...
Page 2: ......
Page 8: ......
Page 12: ...Thermo Electron Corporation iv ...
Page 60: ...Using General Tests Software Thermo Electron Corporation 3 28 ...
Page 70: ...Thermo Electron Corporation 5 4 Connecting Using Accessories ...
Page 74: ...Performing Maintenance Procedures Thermo Electron Corporation 6 4 ...
Page 80: ...Appendix B Parameters Thermo Electron Corporation B 4 ...
Page 88: ...Appendix C Calculations Thermo Electron Corporation C 8 ...
Page 92: ...Index Thermo Electron Corporation I 4 ...
Page 93: ......
Page 94: ...Item No 335904 10001 Rev G 06 03 Component No 335904 10063 Rev G 06 03 ...