Measuring unknowns manually (using Manual 6 or
Single Cell Positioner)
1.
With the
DNA setup
screen displayed, press
Run Test.
2.
Place the blank and unknown in the correct cell
positions. If your instrument is equipped with a
6-Position Cell Holder, you can place up to five
samples in the cell holder.
3.
Press
Measure Blank
to measure the blank. If
your instrument is equipped with a 6-Position
Cell Holder, it automatically moves to the B
position to measure the blank. When it
completes the measurement, it returns to its
previous cell position.
4.
Press
Measure Sample
to start the
measurement. If your instrument is equipped
with a 6-Position Cell Holder, press the cell
position buttons on the keypad to re-position
the cell holder and measure the rest of the
unknowns manually.
DNA with Scan (260/280) and DNA with Scan
(260/230)
The DNA measurement with scan tests include two
test that function almost identically - the only
difference is in the wavelengths used for the
measurements. In both cases the scan is
measured from 225 to 325nm. One test measures
absorbance at 260 and 280nm, while the other
measures absorbance at 260 and 230nm. Refer to
Appendix B for a description of the parameters and
Appendix C for the default values.
To get started, with the BioMate Tests screen
displayed, move the arrow keys to highlight
Nucleic Acid Tests
and press
ENTER
. A list of
nucleic acid test appears. Move the arrow keys to
highlight
DNA with Scan (260/280)
. The
DNA with
Scan (260/280)
parameter screen appears.
Note:
The following screens show the
parameters for the DNA (260/280) test.
For the DNA (260/230) test, Wavelength 2
is set to 230nm.
Note:
If
Cell Correction
is
ON
, you must run the
Setup Correction
program before you can
access the
Run Test
or
Measure
Samples
keys.
Setting up test parameters
1.
With the
DNA with Scan (260/280)
or
DNA
with Scan (260/230)
screen displayed, use the
arrow keys to highlight the name of the
parameter you want to set.
2.
When the parameters are set, you can press
Save Test
to save the test or
Run Test
to
measure the blank and unknowns.
Collecting a baseline scan
Note:
If your instrument is equipped with a 6-
Position Cell Holder, be sure to place the
blank in the B position. The instrument
always uses the B position to collect the
baseline.
1.
With the
DNA with Scan (260/280)
or
DNA
with Scan (260/230)
screen displayed, press
Run Test.
The
DNA with Scan
measurement
screen appears.
Using “Biotests” Software
Thermo Electron Corporation
2-4
Summary of Contents for 335904
Page 1: ...BioMate 3 Series Spectrophotometers Operator s Manual ...
Page 2: ......
Page 8: ......
Page 12: ...Thermo Electron Corporation iv ...
Page 60: ...Using General Tests Software Thermo Electron Corporation 3 28 ...
Page 70: ...Thermo Electron Corporation 5 4 Connecting Using Accessories ...
Page 74: ...Performing Maintenance Procedures Thermo Electron Corporation 6 4 ...
Page 80: ...Appendix B Parameters Thermo Electron Corporation B 4 ...
Page 88: ...Appendix C Calculations Thermo Electron Corporation C 8 ...
Page 92: ...Index Thermo Electron Corporation I 4 ...
Page 93: ......
Page 94: ...Item No 335904 10001 Rev G 06 03 Component No 335904 10063 Rev G 06 03 ...