Save a test
•
View the list of stored tests
•
Measure a blank
•
Measure unknowns
Setting up test parameters
1.
With the
Multiple Wavelengths
screen
displayed, use the arrow keys to highlight the
name of the parameter you want to set.
*Refer to the procedures below for specific
instructions on adding or deleting wavelengths
and factors.
2.
If you have previously selected the
wavelengths you wish to measure, you can
press
Save Test
to save the test or
Run Test
to measure a blank or unknowns.
3.
If you have not selected the wavelengths, you
can add wavelengths and factors as shown
below.
Adding wavelengths & factors
Note:
You can enter factors only when the
measurement mode is set to
Concentration/Factor.
1.
With the
Multiple Wavelengths
screen
displayed, press
Set nms
. The wavelength
screen appears, listing the wavelengths and
factors specified for the measurements.
2.
Use the arrow keys to highlight the position
where you want to enter the first wavelength
and factor pair.
3.
Press
Add nm
.
4.
Enter the values for the wavelength and factor
and press
ENTER
.
5.
When the values are correct, press
Add nm
.
6.
Continue until you have entered all the
wavelengths and factors.
Deleting wavelengths & factors
1.
With the
Multiple Wavelengths
screen
displayed, use the arrow keys to highlight Set
nms. The wavelength screen appears, listing
the wavelengths and factors specified for the
measurements.
Note:
If no wavelength values have been
entered, the wavelength and factor
columns will be empty.
2.
Use the arrow keys to highlight the first
wavelength and factor pair you want to delete.
3.
Press
Delete nm
.
Taking measurements
Taking measurements automatically (Auto 6 or
Auto 3)
1.
The
Multiple Wavelength
acquisition can be
accessed from either the
Set nms
screen
shown above or from the
Multiple Wavelength
set up screen. Press
Run Test
. The
Multiple
Wavelength
measurement screen appears.
2.
Place the blank and the unknowns in the
correct cell positions.
3.
Press
ENTER
to start the measurement. The
instrument automatically measures the blank
first, then measures the unknowns and
displays the sample measurements on the
screen.
Using “General Tests” Software
Thermo Electron Corporation
3-26
Summary of Contents for 335904
Page 1: ...BioMate 3 Series Spectrophotometers Operator s Manual ...
Page 2: ......
Page 8: ......
Page 12: ...Thermo Electron Corporation iv ...
Page 60: ...Using General Tests Software Thermo Electron Corporation 3 28 ...
Page 70: ...Thermo Electron Corporation 5 4 Connecting Using Accessories ...
Page 74: ...Performing Maintenance Procedures Thermo Electron Corporation 6 4 ...
Page 80: ...Appendix B Parameters Thermo Electron Corporation B 4 ...
Page 88: ...Appendix C Calculations Thermo Electron Corporation C 8 ...
Page 92: ...Index Thermo Electron Corporation I 4 ...
Page 93: ......
Page 94: ...Item No 335904 10001 Rev G 06 03 Component No 335904 10063 Rev G 06 03 ...