Recalling a standard curve
1.
Press
Stored Tests
to display a list of stored
tests.
2.
Highlight the standard curve you want to recall.
3.
Press
ENTER
to load the selected standard
curve.
Setting the parameters for a standard curve
1.
Place the standards in the correct cell positions.
2.
Set the parameters for measuring the
standards. Refer to the list of parameters in
Appendix B for a description of each.
•
Enter the Test Name, Wavelength,
Reference Wavelength Correction and
Reference Wavelength.
•
Select the Curve Fit, Units and Sample
Positioner.
•
Set the Number of Standards and Number
of Samples.
•
Enter the low and high limits.
•
Select the settings for the Statistics and
AutoPrint functions.
•
Run the Cell Correction program
Measuring the standards for a standard curve
Measuring standards automatically (using Auto 6 or
Auto 3)
1.
Place the blank and standards in the correct
cell positions.
2.
When all the parameters are correct, press
Run Standards
to set up and run the
standards.
3.
Enter the concentration value into the
Entry
concentration field and press
ENTER
. When
you've entered all the concentration values, the
Standards
screen appears.
4.
Press
Measure Standards
to measure the
blank and standards. The instrument
automatically measures the blank first, then
measures the standards. When the instrument
has measured all the standards, the
Standards
screen appears, showing the
absorbance of each standard, along with the
slope, intercept and correlation coefficient of
the standard curve.
Measuring standards manually (using Manual 6 or
Single Cell Platform)
1.
Place the blank and standards in the correct
cell positions.
2.
When all the parameters are correct, press
Run Standards
to set up and run the
standards.
3.
Enter the concentration value into the
Entry
concentration field. Press
ENTER
4.
When you have entered all the concentration
values, the
Standards
screen appears.
5.
Press
Measure Blank
to measure the blank. If
your instrument is equipped with a 6-Position
Cell Holder, it automatically moves to the B
position to measure the blank. When it
completes the measurement, it returns to its
previous cell position.
6.
Press
Measure Standards
to measure the
standards. If your instrument is equipped with a
6-Position Cell Holder, press the cell position
buttons on the keypad to re-position the cell
Using “General Tests” Software
Thermo Electron Corporation
3-12
Summary of Contents for 335904
Page 1: ...BioMate 3 Series Spectrophotometers Operator s Manual ...
Page 2: ......
Page 8: ......
Page 12: ...Thermo Electron Corporation iv ...
Page 60: ...Using General Tests Software Thermo Electron Corporation 3 28 ...
Page 70: ...Thermo Electron Corporation 5 4 Connecting Using Accessories ...
Page 74: ...Performing Maintenance Procedures Thermo Electron Corporation 6 4 ...
Page 80: ...Appendix B Parameters Thermo Electron Corporation B 4 ...
Page 88: ...Appendix C Calculations Thermo Electron Corporation C 8 ...
Page 92: ...Index Thermo Electron Corporation I 4 ...
Page 93: ......
Page 94: ...Item No 335904 10001 Rev G 06 03 Component No 335904 10063 Rev G 06 03 ...