3
•
Installation
Solvent Monitor
Installation Guide
Page 19
Pump
type
Port 1
Port 2
Port 3
Port 4
Port 5
Port 6
Port 7
Port 8
Quater-
nary
A
B
C
D
-
-
-
Waste
Binary
A1
A2
A3
-
B1
B2
B3
Waste
Dual
A
L
B
L
C
L
-
A
R
B
R
C
R
Waste
Iso-
cratic
A
-
-
-
-
-
-
Waste
16. To check proper connection, pull slightly on the line. If the line does
not slip out of the fitting, it is fixed correctly.
17. To remove any air bubbles from the solvent lines, purge the pump.
The image shows one set of correctly fixed solvent line, solvent
monitoring line, line label, and line sleeve.
Figure 11: Connected solvent monitor line and solvent line fixed with a
line sleeve
3.6.2
Long Solvent Monitor Line
Into each monitored waste container, lead only:
•
The waste of modules of one Vanquish system containing one pump
(one dual pump is supported) and one autosampler and that are
attached to the solvent monitor
•
Small amounts of autosampler needle wash waste liquids