5
•
Maintenance and Service
Page 32
Solvent Monitor
Installation Guide
5.4
Replacing the End Caps
When
•
End cap is:
¨
Contaminated
¨
Discolored or otherwise damaged
•
Semiannually
Parts required
Replacement end cap
Preparations
1. Turn off the pump flow and wait until the system pressure is down
to zero.
Follow these steps
1. Unscrew the cap of the solvent reservoir/waste container and
remove the cap and the solvent monitor line from the solvent
reservoir/waste container.
2. Remove the end cap from the solvent monitor line.
3. Attach a new end cap to the solvent monitor line.
4. Insert the solvent monitor line with the end cap into the solvent
reservoir/waste container ensuring that the end cap rests on the
bottom of the solvent reservoir/waste container and that the
dipstick is straight.
5. To remove any air bubbles from the solvent lines, purge the pump.