Installation
Solvent Monitor
Installation Guide
Page 11
NOTICE
•
Never use defective communication cables. If you suspect that a
cable is defective, replace the cable.
•
To ensure trouble-free operation, use only the cables provided by
Thermo Fisher Scientific for connecting the device.
Parts required
The following parts are needed:
•
From the ship kit:
¨
Cable cover
¨
System interlink cable
¨
USB cable
•
Pair of scissors for cutting the cable cover
•
System interlink cables
Follow these steps
1. Connect the "B" connector of the USB cable to the port labeled
USB 2.0 of the solvent monitor.
2. Connect the "A" connector of the USB cable directly to a USB port of
the data system computer.
If the system is powered, the solvent monitor now powers up. On
the right side of the solvent monitor, an LED shows the status of the
solvent monitor.
NOTICE—Communication problems
Connecting the USB port of the solvent monitor to a USB port of a
module in the system stack may cause communication problems.
Connect the USB port of the solvent monitor directly to the data system
computer.
3. Connect one end of the system interlink cable to the port labeled
System Interlink on the solvent monitor.
4. Connect the free end of the system interlink cable to a free system
interlink port of one of the devices in the system, preferably that of
the detector or the autosampler.