Thermo Scientific
TSQ Altis, Quantis, and Fortis Hardware Manual
129
C
Divert/Inject Valve
You can plumb the divert/inject valve as a loop injector for flow injection analysis or as a
divert valve.
• About the Divert/Inject Valve
• Divert/Inject Valve Configurations
• Divert/Inject Valve Controls
About the Divert/Inject Valve
The external Rheodyne™ MX Series II™ divert/inject valve is a 6-port motorized valve that
switches between two positions. The ports use the standard 10-32 fitting for high pressure
and 1/16 in. OD tubing.
shows the internal flow paths for both positions.
• In the first position, port 1 connects internally to port 2, port 3 connects to port 4, and
port 5 connects to port 6.
• In the second position, the valve rotates clockwise one position so that port 1 connects
internally to port 6, port 2 connects to port 3, and port 4 connects to port 5.
The Method Editor application identifies the valve’s two positions as “1–2” (port 1 to 2) and
“1–6” (port 1 to 6).