22
Maintaining the RF Lens, Exit Lens, MP00 RF Lens, and Lens L0
Removing the Exit Lens, MP00, and Lens L0 (Fortis)
Thermo Scientific
TSQ Altis, Quantis, and Fortis Hardware Manual
99
Removing the Exit Lens, MP00, and Lens L0 (Fortis)
1. Remove the API source interface, see
2. Wearing clean, lint- and powder-free gloves, use a 2-mm Allen wrench to loosen and
extend the two captive Allen screws that secure multipole MP00, and lens L0 to the cage.
3. Pull the multipole MP00 and lens L0 assembly off of the API source interface cage. See
. Place them on a clean, lint-free surface.
Figure 50.
Removing the multipole MP00 and lens L0 assembly (TSQ Fortis)
5. Grasp the two Allen screws and carefully pull the exit lens straight out of the API source
interface cage.
No.
Description
No.
Description
1.
API source interface cage
2.
Allen screws loosened and extended
3.
Lens L0
4.
Multipole MP00 and Lens L0 assembly
4
3
1
2